Medical-Use Withstand Voltage Tester 
Overview

The medical-use withstand voltage tester has all the functions of the conventional withstand voltage tester, and has added an arc (flashover) detection function as per IEC60601-1

Medical-Use Withstand Voltage Tester 

1.Introduction

The RK2672CY medical withstand voltage tester has all the functions of the conventional withstand voltage tester, and has added an arc (flashover) detection function. Through an external oscilloscope, it can detect the “flashover” phenomenon intuitively, accurately, quickly and reliably. When the electrical equipment under test has no “flashover” phenomenon, the oscilloscope will display a stable “Lissajous figure (that is, a closed circle)”. If the electrical equipment under test has a “flashover” phenomenon, the edge of the Lissajous waveform will appear jagged, indicating that there is a “flashover” inside the electrical equipment under test.
This product is also suitable for all kinds of electrical equipment that need to detect “flashover” defects. It is an indispensable withstand voltage tester for medical electrical equipment manufacturers, maintenance departments, users, product quality inspection departments, and technical supervision departments.

 

2. Application

1). Medical equipment: all kinds of new medical equipment and medical supporting equipment, heart monitoring, medical imaging, biochemical analysis equipment, sphygmomanometer and thermometer and other home medical equipment
2). Diagnosis and treatment equipment: X-ray diagnosis and inspection equipment, ultrasonic diagnosis, nuclear medicine, endoscope system, ENT medical equipment, dynamic analysis and medical equipment, cryogenic refrigeration equipment, dialysis medical equipment, first aid equipment
3). Hospital nursing equipment and appliances: all kinds of beds, cabinets, operating chairs, beds, etc.
4). Auxiliary equipment: medical nursing data and image processing equipment, rehabilitation equipment, special equipment for the disabled, etc.
5). Dental medical equipment: dental diagnostic medical equipment, dental surgical equipment, dental technical equipment, medical magnetic resonance equipment

 

3. Specification

ModelRK2672CY
ACWOutput voltage range(0.00-5.00)KV
Maximum output power500VA (5.0KV, 100mA)
Maximum rated current100mA
Current gear2mA, 20mA, 100mA
Output waveformSine-wave
Output waveform distortion≤5% (no-load or pure resistive load)
Test time0.0s-999s 0=continuous testing
DCWOutput voltage range(0.00-5.00)KV
Maximum output power100VA (5.0KV, 20mA)
Maximum rated current20mA
Current gear2mA, 20mA
Ripple factor≤5%
Test time0.0s-999s 0=continuous testing
VoltmeterRange(0.00-5.00)KV
Accuracy±(5%+3 digits)
Resolution10V
Display valueRMS
AmmeterRangeRange 1: 0.1mA~2mA, range 2: 2mA~20mA, range 3: 20mA~100mA
Accuracy2mA: 1μA, 20mA: 10mA; 100mA: 0.1mA
Resolution±(5%+3 digits)
TimerRange0.0s-999s
Resolution0.1St
Accuracy±(1%+50ms)
PLC interfaceOptional
External interfaceYes
DimensionD468mm*H196mm*W378mm
Weight20.65kg
AccessoriesPower cable, HV test wire, ground wire. BNC wire, HV probe
Scroll to Top