Electrical Fast Transient / Burst Generator
Overview

Electrical fast transient/ burst immunity test is a basice EMC test, designed according to IEC61000-4-4, it is to evaluate the impact on the original performance of the device under test when it suffers from transient disturbances.

Electrical Fast Transient / Burst Generator

1. Introduction

Electrical fast transient/ burst is usually be shorted as EFT/B, the EFT/B generator is to simulate the disturbance caused by switching inductive loads by mechanical switches in the power grid, and the transient disturbances are applied between the power line and the ground by the coupling and decoupling network, so the essence of this test is a common mode test, the test levels are given by table 1 of standard IEC61000-4-4.

The test voltage of burst generator is from 0.25KV to 5KV, which can cover all the test levels, and user can define the test level. The high amplitude, short rise time, high repetition rate are completely meet the requirement of IEC61000-4-4. And built-in 380V/16A coupling and decoupling network. The test voltage and CDN can be customized as per the user’s demands.

 

2. Specification

ModelEFT-5K16EFT-5K30
Control and display mode

LCD+touch screen

open circuit output voltage

0.25-5kV ±5%

50Ω load output voltage

0.125-2.5kV ±5%

Wave sharp of the pulse

5/50ns into 50Ω/1000Ω loads

Rise time (tr)

5ns ±30% into 50Ω load

5ns ±30% into 1000Ω load

Duration (td)

50ns ±30% into 50Ω load

50ns -15ns  to 100 ns  into 1000Ω load

Output Impedance

50Ω ±20%

Output polarity

Negative, positive, Positive and negative cycling, positive and negative alternating, positive first then negative, first negative then positive

Repetition frequency

0.1~1200.0kHz ±10%

burst durationStandard

15ms@2.5kHz/5kHz,750μs@100kHz,250μs@300kHz

Adjustable

Nos of pulse can be selected at the range of 1 to 255

Burst duration= (pulse number-1)/repetition frequency

Burst period

100 – 999ms

Burst mode

Synchronous or asynchronous

Synchronous phase

Free set at 0-3600 , resolution: 1°

Test Procedure

Built in IEC standard test procedure or user-define test procedure (1000 groups)

Built-in CDN

(3 phase 5 wires)

AC3-phase 380V/16A, 50/60Hz3 phase 380V/30A, 50/60Hz
DC380V/16A (Max)380V/30A (Max)
Coupling mode

L1,L2, L3, N,PE random combination

Coupling capacitor

33nF ±10%

Environment requirements

Temperature:15°C-35°C; Humidity: 10%-75%

Power requirements

AC220±10%,50/60Hz, 300W

Dimension

470mm(W)×215mm(H)×500mm(depth)

Weight

About 25kg