Ring Wave Generator
Overview

The ring wave generator is designed according to IEC61000-4-12, it is applied to assess the ring wave immunity of electrical and electronic equipment intended for residential, commercial and industrial applications, and power stations and substations

Ring Wave Generator

1.Introduction

The ring wave is a typical oscillatory transient, induced in low-voltage cables due to switching of electrical networks and reactive loads, fault and insulation breakdown of power supply circuits or lightning. This is the most diffused phenomenon occurring in power supply networks, control and signal lines. The ring wave generator is designed according to IEC61000-4-12, it is applied to assess the ring wave immunity of electrical and electronic equipment intended for residential, commercial and industrial applications, and power stations and substations.

 

2. Specification

ModelRWG-4K16
Control and display modeLCD+touch screen
Voltage Oscillation frequency (open-circuit)100kHz±10%
Output voltage(open-circuit)(250V~4000V)±10%
Voltage rise time (open circuit, T1, 10%~90%)0.5μs±30%
Current rise time (short circuit, T2, 10%~90%)≤0.9μs
Open-circuit Voltage Decay Rate40%<(PK2) /(PK1)<110%
40%<(PK3)/(PK2)<80%
40%<(PK4)/(PK3)<80%
Short circuit current at 4000V(PK1)333A±10% @12Ω
133A±10%@30Ω
Interval1s~99s
Number of Impulse1~9999 times
Output modePositive, negative, positive/negative auto
Output Impedance12Ω,30Ω±20%
Phase injection modeasynchronous, synchronous (0-360°)
Coupling and decoupling network(CDN)Single phase 300V/16A or 20A