IEC60601-1 Test Probe Kits
1. Introduction
The test probe kits are meet the requirement of IEC60601-1, includes standard test finger, test pin, straight unjointed test finger, test hook and metal test rod, they are applied to verify the accessibility of living or mechanical hazardous parts and openings in enclosure of medical electrical equipment.
2. Specification
Name | Figure number | Parameter |
Standard test finger (test probe B) | IEC60601-1 clause 5.9.2.1 and figure 6 IEC60950-1 figure 2A IEC61032 figure 2 | Purpose:for identifying the accessible parts of ME EQUIPMENT Specification: 1). Diameter of dactylogryposis: Φ12mm 2). Length of dactylogryposis: 80mm 3). Diameter of baffle. Φ50mm 4). Length of baffle: 100mm |
Straight unjointed test finger with 30N force | IEC60601-1 clause 5.9.2.1 IEC60950-1 clause 2.1.1.1 | Purpose: for verifying the openings preventing the entry of the standard test finger Specification: the size is as same as test probe B, but without joints, 30N force is applied. |
Test Hook | IEC60601-1 figure 7 IEC60065 clause 9.1.7 and figure 4 | Purpose: for verifying the openings of ME EQUIPMENT in question Specification: 1). Length: 180mm 2). Width: 8mm 3). Length of hook: 5mm 4). Thickness: 1mm 5). Material: SUS304 stainless steel |
Test Rod | IEC60601-1 clause 8.4.2 d) | Purpose: for verifying the opening of internal parts that can be touched Specification: 1). diameter: 4mm 2). Length: 100mm 3). Material: metal rod, nylon handle 4). Recommend force applied: 0N, 10N |
Test pin (test probe 13) | IEC60601-1 clause 8.4.2 d) and figure 8 IEC60950-1 figure 2B IEC61032 figure 9 | Purpose: for verifying the opening of internal parts that can be touched Specification: 1). Diameter: Φ3mm (head), Φ4mm (tails) 2). Length of probe: 15mm 3). Diameter of baffle: Φ25mm 4). Thickness of baffle: 4mm |
Ball Pressure Test Apparatus | IEC60601-1 clause 8.8.4.1 and figure 21 | Purpose: for verifying the insulation of ME EQUIPMENT resistance to heat Specification: 1). Ball diameter: 5mm (R2.5mm) 2). Test Pressure: 20N±0.2N 3). Material: Stainless Steel 4). Sample supportor: Φ50mm-length 100mm, cylinder sharp (configure with a hole to insert thermocouple) |