Screwed Glands Torque Test Apparatus
1.Introduction
The screwed glands torque test apparatus can meet the test requirements of IEC60598-1 clause 4.12.5, IEC60884-1, GB2099.1 clause 24.6 and GB7000.1 clause 4.12.5, for testing the screwed glands against the torque strength. Screwed glands shall fit with a cylindrical metal rod having a diameter equal to the nearest whole number of millimeters below the internal diameter of the packing. The glands shall then be tightened by means of a suitable spanner, apply a proper force at a distance of 250 mm from the axis of the screwed gland on the spanner for 1 minute, the luminaire and the glands shall show no damage.
2. Specification
Diameter of metal rods | Φ5*1, Φ7*1, Φ9*1, Φ10*1, Φ12*1, Φ14*1, Φ16*1, Φ18*1, Φ20*1, Φ22*1, Φ25*1, Φ30*1, Φ32*1, Φ40*1 |
Weight | 20N*1, 40N*1, 50N*1, 60N*1 |
Disk diameter | 250mm |
Test time | 60s |
Torque test on glands:
Diameter of test rod (mm) | Metal glands (N.m) | moulded plastic glands (N.m) |
Up to 7 | 6.25 | 2.5 |
7~14 | 6.25 | 3.75 |
14~20 | 7.5 | 5 |
Over 20 | 10 | 7.5 |