IPX1 to IPX6 Comprehensive Waterproof Test Chamber
Overview

The IPX1 to IPX6 comprehensive waterproof test chamber combines the drip box, oscillating tube and jet hose nozzles in one system, it meets the test requirement of IEC60529 regarding the IP level IPX1, IPX2, IPX3, IPX4, IPX5 and IPX6

IPX1 to IPX6 Comprehensive Waterproof Test Chamber

1.Introduction

IPX1-6 waterproof test chamber combines the drip box, oscillating tube and jet hose nozzles in one system. The test chamber can meet the test requirements of IEC60529 regarding the IP level IPX1, IPX2, IPX3, IPX4, IPX5 and IPX6. Waterproof test simulates the condition of vertical water drops, water spraying/splashing, water jets and powerful water jets. The tests are mainly applied to electronic products, luminaire, electric components, motor and house-holding appliance. The waterproof test can help to judge the products’ performance, and provide the basis for design, improvement and inspection.

The test chamber mainly consists by test tunsel and test chamber. The jet nozzles are placed in the test tunsel. And drip box, oscillating tube and turntable are installed on the test chamber. The jet nozzles are for IPX5 and IPX6 testing. The drip box is for IPX1 and IPX2 testing. The oscillating tube is for IPX3 and IPX4 testing.  The test chamber includes water tank and with ball valve to control water level, the water can be recycled. The system is controlled by Panasonic PLC and 7 inch touch screen, user is easy to operate the tests.

 

2. Specification

ModelIPX16B
Control System7 inch touch screen +PLC
Inner Chamber Dimension1300*1300*1450 (W*D*H)
Outer Dimension1750*3800*2150mm (L*D*H)
Diameter of Turntable (mm)Ø600
Bearing capacity50KG
Speed of Turntable1-5r/min
Height of Turntable400-700mm
Tilted angle00 or 150 (in horizontal direction)
Rotation Direction of TurntableForward rotation / forward and reverse alternative / intermittent rotation
Flowmeter adjustment methodManual
Test time0-999min
WeightAbout 380KG
IPX1-2 Drip box
Drip board size800mm*800mm (square)
Aperture of nozzleΦ0.4mm
Nozzles spacing20*20mm
Delivery rate1mm/min(IPX1), 3mm/min(IPX2)
Drip board liftingfixed
IPX3-4 Oscillating tube
Tube Radius (mm)R600
Delivery Rate1.8L/min±5% (IPX3), 2.6L/min±5% (IPX4)
Aperture of nozzles0.4mm
Space between Nozzles50mm
Inner Diameter of Tube15mm
Tube Angle±1200 (IPX3), ±3500 (IPX4) or user-define
Tube Speed4s for 2*1200, 12s for 2*3600
Flow Control ModeManual control
IPX5-6 Jet Nozzles
Aperture diameter of IPX5 nozzle6.3mm
Aperture diameter of IPX6 nozzle12.5mm
Delivery rate for IPX512.5±0.625 L/min
Delivery rate for IPX6100±5 L/min
Water pressure (Mpa)Adjustable as per water flow
Distance from specimen2.5m (distance between nozzles to center of turntable)
Nozzle Swing ModeUP-Down swing (angle, speed adjustable)
Flow Control ModeManual control
ProtectionElectric leakage, hydropenia, short circuit
Working Power380V /50Hz (3 phase 4 wires+ protective ground)
Power supply for specimen220V/50Hz
Power consumption3KW
StandardIEC60529:1989+A1:1999, GB4208-2008, GB7000.1(IPX1, IPX2, IPX3, IPX4, IPX5, IPX6)
Working conditions5℃~35℃ (temperature), ≤85% (R.H.), 86kPa~106kPa (pressure)