UL94 Horizontal and Vertical Flame Test Apparatus
Overview

Horizontal and vertical flame test apparatus is applied to simulate the fire at the early stage surround the electric and electronics appliance, and then evaluate the degree of fire hazard. It is designed according to UL94, IEC60695-11-4, IEC60695-11-3, IEC60950, IEC707, IEC60695-2-2, ISO1210 and etc.

UL94 Horizontal and Vertical Flame Test Apparatus (Model: HVR-4)

1.Introduction

Horizontal and vertical flame test apparatus an meet the test requirements of UL94, IEC60695-11-4, IEC60695-11-3, IEC60950, IEC707, IEC60695-2-2, ISO1210 and etc.. The flame test is to simulate the fire at the early stage surround the electric and electronics appliance, and then evaluate the degree of fire hazard. Horizontal and vertical flame test apparatus is applied to 50W vertical flame test includes V-0, V-1, V-2, horizontal flame tester HB. And 500W vertical flame test includes 5VA, 5VB, HF-1, HF-2 and HBF.

The horizontal and vertical flame test apparatus is controlled by single chip and push button. And the test apparatus is built-in air exhaust, the exhaust system can be preset to delay 1-3mins after power off. The front panel is transparent for easy observation of the experimental process. Fully independent electrical circuit and gas circuit design for safer use.

 

2. Specification

Control unitssingle chip and push button
BurnerΦ9.5mm ± 0.3mm, 100mm length bunsen burner, source: 50W and 500W
Dip angle00, 200, 450 (manual adjustment)
Flame height20mm ± 2mm to 175mm±1mm (adjustable), record by height gauge
Flame time0-999.9s±0.1s (adjustable)
After flame time0-999.9s±0.1s (auto record, manual pause)
After glow time0-999.9s±0.1s (auto record, manual pause)
Number of flame0~9999 times (presettable)
Fuel gas98% methane or 37MJ/m3±1MJ/M3 natural gas
Flow and pressure recordDouble flowmeters and manometer
Gas Pressure0.1Mpa
Gas Flow50W: 105±5ml/min or 18±1 mm,  500W: 965±30ml/min or 117±3 mm
Flame calibrations

1). Flame temperature required: heat up from 1000C±50C to 7000C±30C within the time of 44s±2s (50W) and 54s±2s (500W)

2). Flame height: 50W: 20±2mm (50W)  500W: 40±2mm (inner), 125±10mm (outer)

3). Thermocouple: Φ0.5mm K type thermocouple (The insulating sleeve heat resist to 1250 ℃), 2 pieces to detect the temperature

4). Temperature measuring copper blocks

Test backgroundblack
Volume of chamber0.75 CBM, exhaust hole: Φ70mm
Thermal radiation baffleAutomatic masking
Safe operation interfaceautomatically power off upon open the glass window
AccessoriesPick-up tray, sample clamp, wrapping tissue 12g/m2 ~30g/m2, steel wire mesh, height gauge, Cotton and etc.
Dimension of apparatus1100mm(W)*700mm(D)*1300mm (H)
Power supplyAC220V, 50Hz, ≤500W