IEC60601-1 Medical Electrical Equipment Test Solution

The IEC60601-1 standard is formulated by the International Electrotechnical Commission (IEC). It applies to the basic safety and essential performance of medical electrical equipment and medical electrical systems. The IEC 60601-1 standard, as the core specification in the field of medical electrical equipment, serves like a solid lighthouse, guiding the design, production, and use of equipment.

 

With the continuous advancement of medical technology, medical electrical equipment has become increasingly complex and widely used, and various safety accidents caused by equipment have occurred from time to time. To ensure the safety of patients and operators and standardize the market, the IEC has gradually improved this standard based on comprehensive professional opinions and practical experiences from multiple parties. From the initial version to the continuous revisions and updates today, it has been constantly adapting to emerging technologies and safety challenges.

 

The IEC60601-1 standard involves many testing equipment. The flowing table lists the testing equipment corresponding to the relevant clauses.  

IEC60601-1: 2020 clause and figures

Corresponding to Pego’s testing equipment

Clause 5.7 Humidity preconditioning treatment

High Temperature and Humidity Test Chamber

Clause 5.9.2 Accessible parts

Standard Test Finger (Clause 5.9.2.1, clause 8.9.4, clause 8.10.3, clause 8.11.1, Figure 6)

Unjointed Test Finger (Clause 5.9.2.1)

Test Hook (Clause 5.9.2.2 and clause 8.9.4, figure 7)

 

Clause 6.3 Protection against harmful ingress of water or particulate matter

IP1X 50mm Sphere Probe

IP20 12.5mm Sphere Probe

IP3X Test Probe C

IP4X Test Probe D

Clause 8.4.2 Accessible parts and applied parts (d)

Test Pin (Figure 8, test probe 13)

Metal Test Rod with diameter of 4mm and length 100mm (with 10N force)

Clause 8.4.3 ME Equipment intended to be connected do a power source by a plug

Plug Pins Residual Voltage Tester

Clause 8.7 Leakage currents and Patient Auxiliary currents

Medical Leakage Current Tester (Figure 12)

Clause 8.8.3 Dielectric strength and table 6

Medical Withstand Voltage Tester

Clause 8.8.4.1 Mechanical Strength and resistance to heat

Ball Pressure Test Apparatus (Figure 21)

Heating Cabinet

Clause 8.9.1.7

Tracking Index Tester

Clause 8.9.1.15 Creepage distances and air clearances for defibrillation-proof applied parts (Table 12-16)

Creepage Distances Test Card (30 sizes)

Clause 8.11.3.5 Cord anchorage (Table 18)

Cord Anchorage Pull Force and Torque Tester

Clause 8.11.3.6 Cord guards

Supply Cord Flexing Test Apparatus

Clause 9.4.2.1 Instability in transport position

Incline Plane Tester

Clause 9.5.2.4.3 Movement over a threshold

Threshold Obstruction

Clause 9.4.4 Grips and other handing devices

Handle Strength Test Apparatus

Clause 11.1.3 Measurements

Black Test Corner

Clause 11.6.5 Ingress of water or particulate matter into ME equipment and ME systems

IPX1/2 Drip Box

IPX3/4 Oscillating Tube

IPX5/6 Jet Nozzle

IPX7 Immersion Tank

IPX8 Continuous Tank

Clause 13.2.13.3 ME equipment with motors

Cheesecloth

Clause 15.3.2 Push test

30mm Circular Plane

Clause 15.3.3 Impact test

Impact Test Device with a Steel Ball

Clause 15.3.4 Drop test

Drop Impact Test Hardwood Board

Clause 15.3.5 Rough handling test

Ascending and Descending Step Shock Obstruction (a and b)

Door Frame Shock Obstruction (c)

IEC60601-1