The IEC61010-1:2010 standard specifies the safety requirements that are generally applicable to electrical test and measurement equipment, electrical industrial process-control equipment and electrical laboratory equipment.
The purpose of the requirements of IEC61010-1:2010 is to ensure that Hazards to the Operator and the surrounding area are reduced to a tolerable level.
- Electric shock or burn
- Mechanical hazards
- Spread of fire from the equipment
- Excessive temperature
- Effects of fluids and fluid pressure
- Effects of radiation, including lasers sources, and sonic and ultrasonic pressure
- Liberated gases, explosion and implosion
The following table shows the test requirements covered by IEC 61010-1, which are applicable to Pego’s testing equipment.
IEC61010-1: 2010 clause and figures | Corresponding to Pego’s testing equipment |
Clause 6.2.1 Annex B and figure B.1 Test Probe 11 (Rigid test finger) | |
Clause 6.2.2 Figure B.2 Jointed Finger Probe Clause 15.2 Prevention of reactivating | |
Clause 6.2.3 Metal Test Probe with 3mm diameter and 100mm length | PG-TP4 4mm Metal Test Probe |
Clause 6.2.4 Metal Test Pin with 3mm diameter and 100mm length | PG-TP3 3mm Metal Test Probe |
Clause 6.3 Figure A.1 Measuring Circuits for Touch Current | |
Clause 6.3 Figure A.2 Measuring Circuits for Touch Current 2KΩ Non-Inductive Resistor | |
Clause 6.3 Figure A.2 Measuring Circuits for Touch Current 50KΩ Non-Inductive Resistor | |
Clause 6.3 Figure A.3 Measuring Circuits for Touch Current for Electrical Burns at High Frequencies | |
Clause 6.3 Figure A.4 Measuring Circuits for Touch Current for Wet Location | |
Clause 6.7.1.3 CPT Tracking Index Test | |
Clause 6.7.2.1 Clearance and Creepage Distances | |
Clause 6.8.2 Humidity preconditioning | PG-THC Programmable Constant High Temperature and Humidity Test Chamber |
Clause 6.8.3.1 and 6.8.3.2 AC DC Voltage Test | |
Clause 6.8.3.3 Impulse voltage withstand test Clause 14.8 Circuits or components used as Transient overvoltage limiting devices | LSG-255 1.2/50μs, 8/20μs Impulse Voltage Generator (12Ω resistance) |
Clause 6.10.2.2 Cord anchorage push-pull and torque test | |
Clause 7.4 Stability | |
Clause 7.5.2 Handles and grips | |
Clause 7.6 Wall Mounting | PG-WM Simulated Wall |
Clause 8.2.1 Static Test, Ø12mm hard rod with the hemispherical end, 30N force | PG-TPH12 12mm Hemispherical Test Rod |
Clause 8.2.2 Impact Test | PG-IPT Ball Impact Testing Machine IK-USH Universal Spring Hammer |
Clause 8.3.2 Drop test of hand-held equipment and direct plug-in equipment (50mm Thick hardwood board having a density of more than 700kg/m3) | Hardwood board |
Clause 9.3.2 Flammability classification of Constructional requirements Clause 14.7 Flammability classification of printed wiring boards | |
Clause 10.4.2 Temperature measurement of heating equipment | |
Clause 10.5.2 Non-metallic enclosures | |
Clause 10.5.3 Insulating Material figure 14 | |
Clause 11.1 Overflow test | Overflow Test Apparatus |
Clause 11.6 Specially protected equipment |